Meters and Measurement Devices for Exact Consumption Data Acquisition

Meters and Measurement Devices for Exact Consumption Data Acquisition
Request a Quote

Electromagnetic flow meter features compact structure, integrating the flow sensor and calculator. Its calculator is of vertical type, bringing convenience for users to read data even at a distance.

The measuring principle of this electromagnetic flow meter is on the basis of Faraday’s law of electromagnetic induction. It is suitable for measuring the volume flow of conductive fluids in a closed pipe, such as tap water, industrial water, seawater, waste water, different acid, alkaline and saltine solutions, etc. Thus, this flow meter has wider applications in industrial water drainage, environmental protection, water conservancy, iron and steel, petroleum, chemical engineering and other fields.

Technical parameters
Nominal diameter DN15-DN1000
Accuracy class 1.0 class
Rated working pressure 1.6Mpa
Total power consumption 20W
Output current 4-20mA
Flow direction Forward, backward and net flow 
Protection grade IP65
Working temperature -10℃~+60℃ 
Relative humidity 5%~95%
Min. Straight pipe length Upstream straight pipe: 5D; downstream straight pipe: 3D
Connection mode Flange type
Power supply DC24V@≥1A AC220V 50HZ 
Communication interface Rs485
Communication protocol ModBus-RTU
Environmental class A class 
Electrode materials 316
Electrical environment  Satisfy the requirements of CJ128-200.6.12
Features
  • Flexible parameter setting, the flow sensor is interchangeable.
  • High measuring precision: the measurement is not affected by the variation of fluid density, viscosity, temperature, pressure and conductivity.
  • There is no moving parts or choke parts, thus having no pressure loss and of long life span.
  • Basic Controls provides two kinds of lining materials for customers.
  • Electromagnetic flow meter has functions of self-inspecting and self-diagnosis.
  • Electromagnetic flow meter has a function of bidirectional flow measuring, capable of showing the forward and backward flows.
  • This flow meter is of outstanding electromagnetic immunity. This allows it to operate normally even under a condition of 100KA/m magnetic field strength.
Lining materials
Lining material  Name Property Max. Working temperature Applicable fluid Applicable nominal diameter
Rubber Chloroprene rubber Moderate wear resistance, resistant to corrosion of low concentration acid, alkali and base. 80℃ Tap water, industrial water, seawater DN40-1000
Fluoroplastic Polytetrafluoroethylene Stable chemical property, resistant to corrosion of boiling hydrochloric acid, sulfuric acid and concentrated base. 150℃ Highly corrosive acid, alkali and base solution DN15-1000
PN16-electromagnetic flow meter optional models
DN (mm) a (mm) D(mm) Do (mm) n×A
15 150 95 65 4×14
20 150 105 75 4×14
25 150 115 85 4×14
32 200 140 100 4×18
40 200 150 110 4×18
50 200 165 125 4×18
65 200 185 145 4×18
80 200 200 160 8×18
100 250 220 180 8×18
125 250 250 210 8×18
150 300 285 240 8×22
200 350 340 295 12×22
250 400 405 355 12×26
300 450 460 410 12×26
350 500 520 470 16×26
400 600 580 525 16×30
450 600 640 585 20×30
500 600 715 650 20×33
600 600 840 770 20×36
700 700 910 840 24×41
800 800 1025 950 24×41
900 900 1155 1050 28×41
1000 1000 1280 1170 28×48
PN25-electromagnetic flow meter optional models
DN(mm) a(mm) D(mm) Do(mm) n×A
15 150 95 65 4×14
20 150 105 75 4×14
25 150 115 85 4×14
32 200 140 100 4×18
40 200 150 110 4×18
50 200 165 125 4×18
65 200 185 145 8×18
80 200 200 160 8×18
100 250 235 190 8×22
125 250 270 220 8×26
150 300 300 250 8×26
200 350 360 310 12×26
250 400 425 370 12×30
300 450 485 430 16×30
350 500 555 490 16×33
400 600 620 550 16×36
450 600 670 600 20×36
500 600 730 660 20×36
600 600 845 770 20×39
700 700 960 875 24×42
800 800 1085 990 24×48
900 900 1185 1090 28×48
1000 1000 1320 1210 28×55
Related products
Send Message
Other Products
Inquiry
Inquiry