Medical Suction Unit (suction canister, diaphragm valve, pressure regulator)

Medical Suction Unit (suction canister, diaphragm valve, pressure regulator)
Medical Suction Unit (suction canister, diaphragm valve, pressure regulator)
Medical Suction Unit (suction canister, diaphragm valve, pressure regulator)
Request a Quote

The HS-FYSJ-II medical suction unit adds a membrane in the valve of the regulator, which has stable pressure control and can be adjusted according the demand in the range of 0 to the maximum pressure of pipeline. This device has stable pressure, so it suitable for operating rooms and ICUs as well as for fluid waste collection which specifically include urine and sputum of children in common wards.

Parameter
  • Working pressure: 0 ~ -0.075Mpa (adjustable)
  • Gas flow rate of suction: 0 - 30L/min (adjustable)
Features
  • The valve of vacuum pressure regulator that is made of H59-1 brass has longer life.
  • The lid of canister is made from ABS material.
  • The jar of canister is made from PC material and can be sterilized at high temperature and reused.
  • Under the pressure of -0.075Mpa, the adjusted pressure will never change.
  • The function of float ball: when the volume of fluid up to the setting capacity, float ball will seal up the outlet automatically to shut off the flow.
Operating Instructions
  • This type of device is an integral suction unit, can be used with a matched canister of any capacity. After connecting the vacuum tube to the canister, plug the unit into the gas terminal outlet.
  • The jar and plastic lid should be washed with tap water and then dipped in disinfectant to sterilize before use. Inserting the plug of control knob into gas terminal outlet, the plug will be locked when a “click” is heard.
  • Turn on vacuum control knob slowly and show the negative pressure value of central suction; then, according to patient’s requirement for the volume of fluid waste, control the flow rate via fine tuning knob.
  • By the end of utilization, turn off the vacuum control knob and unplug the vacuum connection tube.